If the photograph is clicked, it grows.

Keysight E5061B 3L5/005/1E5 VNA+Impedance Analyzer
JPY3,000,000 1 stock
net422



Source
Frequency Range:5 Hz to 3 GHz
Resolution:1mHz
Source stability:± 7 ppm (5 to 40 °C)
CW accuracy:± 1 ppm ± 1 mHz

Level accuracy
(Option 3L3/3L4/3L5)
± 0.8 dB (at 0 dBm, 50 MHz absolute)
± 1.0 dB (at 5 Hz to 1.5 GHz, 0 dBm,
relative to 50 MHz)
± 1.5 dB (at 1.5 GHz to 3 GHz, 0 dBm,
relative to 50 MHz)

Level lineality
(Option 3L3/3L4/3L5)
± 0.75 dB (at –10 to 10 dBm, 0 dBm
reference)

Output Power Range
(Option 3L3/3L4/3L5)
–45 dBm to 10 dBm
Level resolution:0.05 dB


Input

Absolute amplitude accuracy < ± 3 dB
(at 0 dBm)

Crosstalk
(Option 3L3/3L4/3L5)
–85 dB (at 5 Hz to 100 Hz)
–100 dB (at 100 Hz to 9 kHz)
–110 dB (at 9 k to 100 kHz)
–115 dB (at 100 kHz to 3 GHz)

Noise floor (Option 3L3/3L4/3L5)
10 Hz bandwidth
-90 dBm (100 Hz to 9 kHz)
-100 dBm (9 kHz to 100 kHz)
-105 dBm (100 kHz to 1 MHz)
-110 dBm (1 MHz to 3 GHz)

E5061B-3L5 + 005 Impedance Analyzer function

Impedance parameters:IZI, θz , IYI, θy, Cp, Cs, Lp, Ls, Rp, Rs, D, Q, R, X, G, B

Measurement methods:Port 1 Reflection, Port 2 Reflection, Port 1-2 Series-thru, Port 1-2 Shunt-thru,
Gain-phase Series-thru, Gain-phase Shunt-thru

Equivalent circuit analysis:3-component model (4 models: A/B/C/D), 4-component model (1 model: E)

Test frequency: 5 Hz to 500 MHz/1.5 GHz/3 GHz (Port-1 and 2), 5 Hz to 30 MHz (Gain-Phase test port), 1 mHz resolution
AC source power(Osc level): –45 to + 10 dBm (when the source port is terminated with 50 Ω)
DC voltage bias: 0 to ±40 V (100 mA max.) 1 mV resolution (0 V to ±10 V), 4 mV resolution
(10 V to 40 V, -10 V to -40 V), DC bias monitor available (voltage or current)

Option:
3L5 LF-RF network analyzer with DC bias source, 5 Hz to 3 GHz
1E5 High stability timebase
020 Standard Data Storage
005 Impedance analysis for LF-RF network analyzer
Accessory:Mouse,Keyboard,Manual